![](https://ae01.alicdn.com/kf/Sa0423aadc37648d3a2fd6c7ae48167ce6/UVA-UVB-UV-LS125.jpg) ![](https://ae01.alicdn.com/kf/Sc2689311af89437f869b9a9f3d7177e1n/UVA-UVB-UV-LS125.jpg) ![](https://ae01.alicdn.com/kf/Sb716e4f7767445389d641f37e20e921co/UVA-UVB-UV-LS125.jpg) ![](https://ae01.alicdn.com/kf/Sa216845d03a14a3e9fd614483b99d3e6Z/UVA-UVB-UV-LS125.jpg) aHR0cDovL2ZyZWVzaGlwLmNvLmty![](https://ae01.alicdn.com/kf/Sa0423aadc37648d3a2fd6c7ae48167ce6/UVA-UVB-UV-LS125.jpg) ![](https://ae01.alicdn.com/kf/Sc2689311af89437f869b9a9f3d7177e1n/UVA-UVB-UV-LS125.jpg) ![](https://ae01.alicdn.com/kf/Sb716e4f7767445389d641f37e20e921co/UVA-UVB-UV-LS125.jpg) ![](https://ae01.alicdn.com/kf/Sa216845d03a14a3e9fd614483b99d3e6Z/UVA-UVB-UV-LS125.jpg) ![](https://ae01.alicdn.com/kf/Saae7b2e843934c8fb925f9dea7a4ea2eq/UVA-UVB-UV-LS125.jpg)
- ¸ðµ¨ ¹øÈ£: LS125
- ºê·£µå À̸§: NoEnName_Null
- ±Ù¿ø: CN (Á¤Ç°)
- Model Number: LS125
- Power: Electronic
- Product name: LS125 UV light meter
- Application: Curing, Exposure, Printing, Sterilization industry
- Function: Measure UV power and energy value
- Power supply: 4 AAA alkaline batteries
- Certificate: CE
- Dimension: 148mm * 76mm * 26mm (L*W*H)
- Display: 240*160 dot matrix LCD
- Probe connection: Push pull (Snap-in style),Aviation socket
- Accuracy: ¡¾10%
Overview
LS125 Muiti-probe UV Light Meter (2nd)
LS125 is a smart host that can support 9 different probes. It can intelligently identify different probes. This host is connected to the probe through an aviation plug. The main function of the host is to display the test data. The host equipped with the probe can be used for ultraviolet intensity and energy detection. When using the host, you need to buy the corresponding probe to use with it according to the test needs.
uva uvb meter uv radiometer light irradiance meter ultraviolet light measurement
1. Multi-probe UV light meter. The host and probe are linked by aviation plugs. One host can intelligently match 9 different UV probes to meet different test needs.
2. Digital probe: adopt advanced digital probe technology, the probe is not easy to be interfered and provides excellent measurement accuracy. 3. The instrument can intelligently identify the probe model and the probe will adapt the display interface automatically.
4. A variety of units are available: The instrument can provide different unit selections based on the identified probe model. After connecting the probe, you can choose the appropriate unit in the setting mode. 5. Abundant data statistics function: real-time power value, maximum value, minimum value, average value, measurement time, energy value can be displayed at the same time. (You can also choose not to display the energy value in the setting mode)
6. Data recording: 9 sets of measurement data can be saved. During the measurement, press the "HOLD" to lock the maximum value and the current value during the measurement at the same time. The data will be saved even after the power is turned off. (Except when the battery is removed) 7. Ultra-low power consumption: The instrument is powered by four AAA batteries. The UV measurement device will automatically shut down after 10 minutes without any operation, which can avoid wasting battery power and prolong the service life of the instrument.
According to the spectral response range, application, measuring range and test aperture size of the light source, we have introduced 9 different types of probes (probe must be purchased separately). The selection table is as follows:
1. Probe spectral response range, test aperture and application
No.
|
Probe Type
|
Spectral Response Range
|
Test aperture
|
Application
|
1
|
UVC-X0
|
230-280nm
|
¡Ë10mm
|
254nm UV sterilization mercury lamp
|
2
|
UVCWP-X1
|
230-280nm
|
¡Ë10mm
|
254nm UV sterilization mercury lamp,waterproof
|
3
|
UVCLED-X0
|
240-320nm
|
¡Ë10mm
|
260nm-285nm UV LED germicidal lamp
|
4
|
UVB-X0
|
280-315nm
|
¡Ë10mm
|
General UVB intensity and energy measurement
|
5
|
UVA-X0
|
315-400nm
|
¡Ë10mm
|
High pressure mercury lamp in UV curing industry
|
6
|
UVA-X1
|
315-400nm
|
¡Ë10mm
|
General UVA intensity and energy measurement
|
7
|
UVALED-X0
|
340-420nm
|
¡Ë10mm
|
Area light source of UV LED in UV curing industry
|
8
|
UVALED-X1
|
340-420nm
|
¡Ë1mm
|
Point light source of UV LED in UV curing industry
|
9
|
UVALED-X3
|
340-420nm
|
¡Ë10mm
|
General UVA+UVV LED, low power measuring range
|
2. Probe power measuring range, optional units and size
No.
|
Probe Type
|
Power Measuring Range
|
Optional Units
|
Probe Size
|
1
|
UVC-X0
|
0 - 200000 ¥ìW/cm²
|
¥ìW/cm², mW/cm², W/m²
|
Diameter 39mm* Thickness 15mm
|
2
|
UVCWP-X1
|
0 - 200000 ¥ìW/cm²
|
¥ìW/cm², mW/cm², W/m²
|
Diameter 50mm* Thickness 20mm
|
3
|
UVCLED-X0
|
0 - 200000 ¥ìW/cm²
|
¥ìW/cm², mW/cm², W/m²
|
Diameter 39mm* Thickness 15mm
|
4
|
UVB-X0
|
0 - 200000 ¥ìW/cm²
|
¥ìW/cm², mW/cm², W/m²
|
Diameter 39mm* Thickness 15mm
|
5
|
UVA-X0
|
0 - 2000 mW/cm²
|
mW/cm², W/m²,
|
39mm*32mm*15mm(L*W*H)
|
6
|
UVA-X1
|
0 - 200000 ¥ìW/cm²
|
¥ìW/cm², mW/cm², W/m²
|
39mm*32mm*15mm(L*W*H)
|
7
|
UVALED-X0
|
0 - 20000 mW/cm²
|
mW/cm², W/cm², W/m²
|
39mm*32mm*15mm(L*W*H)
|
8
|
UVALED-X1
|
0 - 20000 mW/cm²
|
mW/cm², W/cm², W/m²
|
39mm*32mm*15mm(L*W*T)
|
9
|
UVALED-X3
|
0 - 200000 ¥ìW/cm²
|
¥ìW/cm², mW/cm², W/m²
|
39mm*32mm*15mm(L*W*H)
|
3. UV light meter host parameters
|
|
|
|
|